SECTION 10 APPENDIX I
_______________________________________________________________________
10-2
10.2 SENSOR SIMULATOR: P/N: 186-0137
Description of device:
The Sensor Simulator is a service tool that allows for the verification of proper impedance
values being detected by the BIS VISTA Monitoring System during the Sensor Check. This
test is part of the initial test that each system performs. The simulator also allows for safety
testing of the system in the field by allowing connection of the test equipment to the
monitor via the Patient Interface Cable (PIC).
NOTE:
The life expectancy of the Sensor Simulator is 25 connect / disconnect cycles.
S1 – S4 connect to 4 input signal pins on the PIC. The Inputs (+ and -) are where one
connects the test signals to test the BIS VISTA system.
Figure 8 - Schematic of Sensor Simulator Circuit
Test types allowed:
Sensor Check:
The Sensor Check tests and verifies that the monitor is reporting the proper impedance
values provided by the Sensor Simulator. This procedure verifies functionality of the BIS
VISTA Monitoring System. See Section 7.2.3 “PIC Checkout Procedure” for instructions.
Safety Testing: Leakage Current
Leakage Current testing should be performed by a qualified Biomedical Engineering
Technician or authorized personnel only. See Section 6.4 “Checking Leakage Current” for
instructions.