SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-5
Status Messages: Causes: Corrective Actions:
16
Last Sensor Check
Failed
1. At least one element of
sensor has too high
impedance, and EXIT
pressed (before sensor
check completes).
2. Incorrect sensor
application.
3. Poor sensor connection.
4. Defective PIC.
5. Defective sensor.
6. Defective BISx.
1. Verify Sensor Check
passes.
2. Read Instructions on
sensor package and re-
apply sensor.
3. Check sensor connection.
4. Replace the PIC.
5. Replace sensor.
6. Replace BISx Bulkhead
Connector or replace BISx.
27
Excessive Artifact
Detected in Signal
The Signal Quality is less
than half of the level
desirable for optimal
monitoring conditions. This
may occur as the result of
artifact such as those
generated from motion or
eyeblinks.
1. Artifact, such as those
generated by motion or
eyeblinks, is causing loss
of EEG recognition.
2. EMG Bar indicates
electrical activity that
may be interfering with
EEG recognition.
3. PIC is defective.
4. BISx is defective.
Note: This message may
occur as the results of
artifact (non-EEG signal)
such as those generated
from motion (patient
movement or eyeblinks) or
the presence of electro-
cautery, warming blankets,
or other devices
1. If ARTIFACT label
appears above the EEG
waveform box, attempt to
identify and eliminate
artifact source.
2. If EMG bar is illuminated,
attempt to determine and
eliminate cause.
3. Verify Sensor Check
passes. If not, replace PIC.
4. Replace BISx Bulkhead
Connector or replace BISx.