Aspect BIS VISTA Spécifications Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 94
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
SECTION 6 PREVENTIVE MAINTENANCE
_______________________________________________________________________
6-4
6.2.3 Patient Interface Cable (PIC) Checkout Procedure
The successful completion of this test will verify function of the BIS VISTA system from
the BISx circuits to the patient connector. Since the conductors used are located both in the
BISx and the PIC, use a swapping technique to isolate the faulty component.
Use a Sensor Simulator or make a Test Sensor for this test. See Section 10.3 “Test Sensor”
for details.
IF ANY FAILURES ARE NOTED, SEE SECTION 7.3, “BIS
VISTA System Messages And Corrective Actions”.
1. Using a known good BIS VISTA monitor (see section 6.2.1 “Monitor Checkout
Procedure) and a known good BISx (see section 6.2.2 “BISx Checkout Procedure)
connect the PIC that you are testing to the BISx, and connect the BISx to the BIS
VISTA monitor.
2. Press the monitor’s ON/Standby button to start up the monitor and BISx. At the
completion of the power up self-test, verify that the screen message “Connect sensor
or cable” displays.
3. Connect a Sensor Simulator (see Section 10.2) or Test Sensor (see Section 10.3) to
the PIC cable.
4. A sensor check is initiated automatically when the sensor and PIC are connected to
the BISx. (It may also be initiated by pressing the [SENSOR CHECK] touch key.)
The message, “Sensor Check in Progress” appears. When the sensor successfully
passes the test, the BIS Trend screen displays.
5. If the Sensor Check is not immediately successful, the Sensor Check Graphic
Screen displays automatically. This screen shows a sensor with each electrode
numbered. Colors indicate the status of each electrode.
White hollow circle – No status is available (Lead is off).
Green circle – The electrode impedance is within the acceptable range.
Red blinking circle – The electrode impedance is not within the acceptable range.
The impedance value for each electrode, in kilo ohms, appears on the screen along
with its status:
PASS - An electrode passes if the impedance for that electrode is less than 7.5
kilo ohms. The ground electrode (element #2) must be less than 30 kilo ohms to
pass.
HIGH - An electrode is labeled “HIGH” if its impedance value is above 7.5 kilo
ohms. As long as the combined impedance of electrodes #1 and #3 and the
combined impedance of electrodes #1 and #4 are less than 15 kilo ohms, the
sensor check will be considered successful.
NOISE - If the signal from the electrode goes beyond the measurable range, the
label “NOISE” displays.
LEAD OFF - If the impedance check indicates that the electrode is not in
contact with the patient, the label “LEAD OFF” displays.
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 93 94

Commentaires sur ces manuels

Pas de commentaire