SECTION 8 SERVICING THE BIS VISTA MONITORING SYSTEM
8-2
8.3 Replacing the Monitor
Parts Required: Tools Required:
BIS VISTA Monitor P/N 185-0151 None
To replace the monitor:
1. Put monitor in Standby mode by pressing the ON/Standby button. The light to the
right of the button should be yellow or off.
2. Unplug the power cable from the wall and remove it from its receptacle in the rear
of the monitor.
3. Disconnect the BISx from the front of the monitor. To do this, grasp the connector
on the BISx Monitor Interface Cable and pull. DO NOT twist or pull on the cable.
4. If necessary, dismount monitor from pole clamp by removing set screw from clamp,
then depressing top of clamp shoe (blue plastic clip at top of aluminum pole clamp)
and slide the monitor up and off of the clamp.
5. To install the replacement monitor, insert power cord into receptacle on rear of
monitor.
6. To install the BISx, hold the cylindrical connector with flat side up (at 12 o’clock
position) and insert firmly into BISx port on front of monitor.
7. Re-mount pole clamp assembly.