SECTION 8 SERVICING THE BIS VISTA MONITORING SYSTEM
8-3
8.4 Replacing the Battery
Parts Required: Tools Required:
Battery Replacement Kit P/N 186-0208 Philips #2 screwdriver
Caution:
All repairs to the BIS VISTA Monitoring System should be made only
by a qualified Biomedical Engineering Technician or other authorized
personnel.
To replace the battery, you will need a Philips #2 screwdriver. Follow the instructions below:
1. Unplug A/C line cord from the monitor.
2. Lay monitor screen-side-down on a scratch-free work surface so that the
Battery/Power Supply cover is accessible.
3. Remove 4 screws from the Battery/Power Supply cover and remove the cover. Note
the position of the battery cable.
4. Squeeze the battery connector latch to disengage it from the back of the monitor and
remove the old battery.
5. Lay the new battery in the recess with the wires at the top, and plug in the connector.
6. Replace the cover and four screws (hand-tighten only) and reconnect the A/C power
cord.