SECTION 3 PRINCIPLES OF OPERATION
_______________________________
_
_______________________________________
3-4
3.2.1.1 BISx Signal Conditioning
The input protection circuits are designed to protect the input from destruction by electric
shock from sources such as electrostatic discharge (ESD) or defibrillation. The protection
circuits also reduce the effects of high frequency ambient noise from sources such as electro-
cautery and other devices.
Input signals are amplified by instrumentation amplifiers, which have a fixed gain. The
amplifiers have DC servos, which remove the signals below high pass cutoff frequency. In
the event of amplifier overload, the servos are changed to a higher frequency to facilitate fast
recovery (blocking) under control of the host processor.
Each channel is further amplified to the level required by the A/D converters. The
amplifiers also serve as filters to prevent aliasing by the converters.
3.2.1.2 BISx Impedance Testing
In the default state of the BISx the combined channel electrodes’ impedance is continuously
checked. A small current (approximately 1 nanoampere) is injected into each electrode at 128
Hz, just above the EEG band. The resulting voltages are measured. Equal but opposite
currents are injected into the (+) and (-) electrodes simultaneously while the digital signal
processor measures the resulting voltage. BIS monitoring is performed while combined
impedance is checked.
The BISx measures the individual electrode impedance during a sensor check by injecting
current into the REF electrode only. Individual electrode impedance is derived by
subtracting the resulting value from the combined value. BIS monitoring is interrupted while
individual impedance is checked.
The ground electrode impedance is also measured while injecting current into the REF
electrode. BIS monitoring is interrupted while the ground impedance is checked. Ground
impedance checking occurs when a sensor check is performed and thereafter on a 10-minute
schedule during patient monitoring.
The impedance check signal can occasionally interfere with other monitoring equipment
connected to the patient. Evoked potential monitors are particularly susceptible because they
use a wide bandwidth. The automatic impedance check feature can be turned off by selecting
“Impedance Checking – OFF.” (See Operating Manual for specific instructions).
3.2.1.3 BISx Processor and Communications Circuits
The BISx contains: an analog to digital (A/D) converter for each channel, the monitor
interface, the sensor interface and the power supply circuits. A crystal controlled BISx
master clock is on this board. This clock is the system’s BIS processing clock.
A/D Conversion
There are two independent sigma-delta modulators for the two channels. These run at 16384
samples per second.