Aspect BIS VISTA Instructions d'exploitation

Naviguer en ligne ou télécharger Instructions d'exploitation pour Matériel d'enregistrement Aspect BIS VISTA. Aspect BIS VISTA Operating instructions Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 61
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Operators Manual
Expert Anesthesiology
Monitoring System
Team 2:
Timothy Morin
Nathan White
Kane Killelea
Sponsored by:
Joseph H. McIsaac and John D. Enderle
March 23, 2007
Client Contact:
Dr. John D. Enderle
Editor-in-Chief, EMB Magazine
Biomedical Engineering Book Series Editor
for Morgan and Claypool Publishers
Program Director & Professor for Biomedical
Engineering
University of Connecticut
Bronwell Building, Room 217C
260 Glenbrook Road
Storrs, Connecticut 06269-2247
(860) 486-5521
Joseph McIsaac
Chief, Trauma Anesthesia
Hartford Hospital
Suite JB300,
Department of Anesthesiology
80 Seymour St.
Hartford, CT 06102
(860)-545-2117
Vue de la page 0
1 2 3 4 5 6 ... 60 61

Résumé du contenu

Page 1 - Monitoring System

Operators Manual Expert Anesthesiology Monitoring System Team 2: Timothy Morin Nathan White Kane Killelea Sponsored by: Joseph H. McIsaac

Page 2

91 Introduction 1.1 General overview of the device The objective of the Anesthesia Vital Signs Monitor is to create a program that not only acts

Page 3

10Anesthesia Vital Signs Monitoring System Figure 1: Anesthesia Vital Signs Monitoring System As shown in Figure 1, most of the parts run ind

Page 4

11 Figure 2: BIS Vista attached to a patient and to the laptop Figure 2 shows the data being transferred from the BIS Aspect monitor to the la

Page 5

121.2 Operating Instructions BIS Monitor and LabVIEW program 1. Power on the BIS monitor. 2. Next the transducer must be plugged into the cable

Page 6 - Features

13 Figure 4: How the transducer is to be worn he sensors should be pressed firmly onto the forehead until the BIS Aspect moni

Page 7 - Table of Contents

14The BIS displays each sensor on the turn on screen and displays which ones are n within an adequate range of SQI to have a low enough value of no

Page 8 - Figures Page

154. Power on the Macbook Pro Figure 7: Power on Macbook Pro laptop 5. Once the Macbook Pro turns on, the LabVIEW executable file then will ju

Page 9

16. Ensure that all connection have been made between the power supply and the circuit. 2. Make sure that the serial cable is connected as shown i

Page 10 - 1 Introduction

17 4. Turn on power supply. As shown in figure 11, the power supply can be turned on by just pressing the on button to the l position. Figure 11:

Page 11

18 • PDA Figure 13: Cingular 8525 Cell phone/PDA 1. Turn on the Cingular 8525 device, found at location 1 in Figure 13. 2. Remove the scree

Page 12

1Important Safety Instructions When operated properly, the Anesthesia vital signs monitor can provide a safe, easy, and convenient way to provide

Page 13

19 Click on “Add New Device” • This should discover your Bluetooth adapter for your laptop computer. • Input “EAMD server” into Display Name •

Page 14

202 Maintenance The maintenance for the a salterations are being made to the over all design of the program. For fast reference the maintenance s

Page 15

21Plug in charger as shown in Figure 15. 3. Figure 15: Plug in charger 4. Plug in other end of charger to the PDA as shown in Figure 16. Figu

Page 16

22. Spray a small amount of glass cleaner or other LCD safe cleaning product on the screen as shown in Figure 17. Cleaning the LCD screens 1 Figur

Page 17

23ed to clean the screen of the PDA. Start at number 1 under verall Maintenance: These methods can also be usthe “Cleaning the LCD section” and s

Page 18

24The maintenance for the LabVIEW programs should only be performed by an nician with adequate knowledge of the program design. The irst step in La

Page 19

25ake sure proper virus software is installed if the laptop is going to be on a live network. einstallation of Software PDA S nk or follow the Lap

Page 20

26 Figure 21: D-Link Driver CD

Page 21

27• Don’t plug in Adapter until the Install software prompts you to. Figures 22 and 23: Installing Bluetooth • Inserting the CD into the CD-RO

Page 22

28 Figures 24, 25, and 26: Installation of Bluetooth. Wait while the program installs Bluetooth Utility Setup

Page 23

2Parts and Accessories All included hardware and necessary equipment for the successful operation of the Anesthesia monitoring device: • MacBook

Page 24

29 Figure 27: Bluetooth Manager Figure 28: Click “OK” Figure 29: Wait

Page 25

30 Figure 30: Status

Page 26

31 Figures 31 and 32: Click “Finish”, then click “Yes” After the Computer reboots it will be Bluetooth enabled and you may proceed to setting up t

Page 27 - Figure 21: D-Link Driver CD

32• Once the manager is opened click on New connection the button looks like the figure below. Figure 34: Creating a Bluetooth connection • Thi

Page 28

33 Figure 36: Select your device’s name and click “Next”. The software will take a few moments to establish the connection. Figure 37: Click “N

Page 29

34• After the Connection is established the Bluetooth manager will return to the initial screen and you will see your newly acquired Virtual serial

Page 30

35 Installing USB/Serial Adapter (BELKIN) Figure 39: Click “Next” to continue install CAUTION: Make sure to read th ent that will come in the n

Page 31

36 Figure 41: Select en click “Finish” ongratulations: have successfully installed the Belkin USB to PDA Adapter Driver Install rocess. This w

Page 32

37 Technical Description The Technical Description is broken down into individual parts for a fast efficient reference. There is the Data Acquisi

Page 33

38er that your computer assigns to it, on the list of connect sing: ( ). This method is mainly used for serial connection testing. Although it can

Page 34

3 • BIS Vista Monitor • LCD and Circuit • PDA with Windows Operating system

Page 35

39 ork at will be sniffed is chosen by highlighting the connection with the blue color on the list of adapters in the program. This program is very

Page 36

40 Figure 46: Block diagram for the Packet Sniffer.vi Next the correct broadcast must be obtained and separated from the rest of the broadcast data

Page 37

41Data packet three: 10/25/2007 16:33:57| 8| 4|On |High |Off | 70|No | 0.0| 22.9| 00f1| 97.3| 55.

Page 38

42y done with the BIS Vista. The Solar 8000i is a much more omplicated piece of software. This software would require an understanding of the T

Page 39

43 Figure 50: Block diagram to retrieve and parse BIS Vista broadcast Another program was written to simulate the data that would be outputted b

Page 40

44 Figure 52: Cingular 8525 Before the testing can began the LabVIEW program ActiveSync has to be installed onto the lab computer which will be

Page 41

45 Figu face ent uld be used because it has a greater ld e s used by the TCP/IP networks. Several measures can be taken to prove the program if a d

Page 42

46 Figure 54: This diagram shows how the device will eventually work with the networks already present at Hartford Hospital Using the RWN provide

Page 43

47This adapter is accompanied by drivers which allowed the personal computer to detect the Bluetooth device. Once the drivers are installed an icon

Page 44

48 to When the initialize button is pressed the Bluetooth manager can establish the connection as Figure 49 shows. Once the connection is created

Page 45

4 • D-Link Bluetooth USB transmitter • BIS Semi Reusable Sensor

Page 46

49 Figure 59: Server VI Microcontroller This device uses the PIC 16F877-20/P microcontroller from Microchip because it satisfies all of the needs

Page 47

50The microprocessor uses a 6 MHz oscillator for timing control that is connected to pins 13 and 14. Pins 19 through 21 are used as digital outputs

Page 48

51 e DB9 is connected to ground. Pins 11 and 12 of the MAX232A are connected to pins 25 and 26 of the PIC, respectively. Pin 14 of the MAX232A is

Page 49

52 F s racters to be displayed properly on the LCD, VEE from pin 12 on the LCD igure 63: LCD pin assignment n order for chaImust be passed through

Page 50

53 Figure 65: Contrast potentiometer IMPORTANT: Do not touch any of the electri product as electric shock can result. If any of t lt a qualified

Page 51

54 Complete Integration Figure 66: 1 The LCD protoboard, 2 LCD, 3 PDA, and 4 the laptop with the front panel Figure 66 shows each of the four p

Page 52

55 Figure 67: BIS Monitor transferring data to the laptop using USB to Serial Converter The complete LabVIEW program is displayed in Figures 68 an

Page 53

56 Figure 68: The front panel for the final program with both the PDA and the BIS monitor LabVIEW programs combined

Page 54

57 Figure 69: Simulated BIS information being parsed and sent to the PDA Bluetooth program The PDA transmission of live data is accurate but th

Page 55

58 Figure 70: PDA displaying real time data, where the peaks represent the level of consciousness of the patient

Page 56

5Features The anesthesia vital signs monitor has the ability to supply the anesthetist with the following information and technology: • Real tim

Page 57

59Electrical Specifications Laptop Operating Voltage ** ** LCD Operating Voltage 5.0 Volts PDA Operating Voltage ** ** Laptop Operating Cu

Page 58

60roubleshooting/Technical Support le se H o Fix It T Problem Possib Cau ow tLaptop does 1. No not turn on powe 1. Plug in the Laptop and turn

Page 59

6Table of Contents 1. Introduction 9-19 1.1 General overview of the device 9-11 1.2 Operating Instructions 12-19 1.

Page 60

7 Figures Page Figure 1: Anesthesia Vital Signs Monitoring System 10 Figure 2: BIS Vista attached to a patient and to the la

Page 61

8Figures Page Figure 45: Front panel of the Packet Sniffer.vi 39 Figure 46: Block diagram for the Packet Sniffer.vi 40 Figure

Commentaires sur ces manuels

Pas de commentaire